As the clock runs out on the 3M™ Novec™ phaseout, many manufacturers are wondering how to switch their vapor degreasers to a 3M™ Novec™ replacement solvent such as BestSolv® Engineered Fluids.
Here are the steps to use when converting your vapor degreaser to a new solvent.
Remove Old Solvent
Completely remove the old solvent from the vapor degreaser before adding the new solvent. Any residual old solvent left in the lines may be flushed with the new solvent. Do NOT use water to clean the vapor degreaser or flush the lines. If residual old solvent remains or has turned acid previously, neutralize the degreaser and all surfaces.
Add New Solvent
Review the Safety Data Sheet for the new solvent prior to first use. As always when handling solvents, use appropriate protective equipment as mentioned in the Safety Data Sheet. Since most solvents are designed to evaporate quickly, be careful not to leave the solvent container open any longer than necessary.
Identify Boiling Point of New Solvent
The boiling point on your new solvent may be different from that of the old solvent. Vapor degreaser settings are based on the boiling point of the solvent used, and should be adjusted when a new solvent is used in the degreaser.
Use the Safety Data Sheet, Technical Data Sheet, or other publicly available source to identify the boiling point of the new solvent. For BestSolv® Engineered Fluids, you can find this information through our Technical Data Sheets page.
Adjust Vapor Degreaser Settings
Use the new solvent’s boiling point as a reference for changing the following settings on your vapor degreaser:
| Vapor Degreaser Setting | Set Point |
| Liquid Temperature Control (LTC) | Boiling Point plus 7-10 °F |
| High Temperature Control (HTC) | Boiling Point plus 10-15 °F |
| Solvent Vapor Control (SVC) | Boiling Point minus 10 °F |
| Vapor Up Control (TH-1) | Boiling Point minus 10 °F |
For example, a solvent with a boiling point of 165 °F would have a Vapor Up Control (TH-1) set point of 155 °F (165 – 10 = 155).
That’s it! Your vapor degreaser is ready to return to precision cleaning with your new solvent.
NOTE: Vapor degreaser controls differ by manufacturer. Some may not contain all of the settings described.
For further information about vapor degreaser settings, see the reference section below.
Vapor Degreaser Settings Reference
- Liquid Temperature Control (LTC): Set 7-10 °F above the boiling point of the solvent being used. A 10 °F rise in the boiling point indicates that the boiling sump has accumulated the equivalent oil solution of approximately 30% by weight, along with any other contaminants. Boil down and addition of fresh solvent is recommended.
- High Temperature Control (HTC): Set 10-15 °F above the boiling point of the solvent being used. A 10 °F rise in the boiling point indicates that the boiling sump has accumulated the equivalent oil solution of approximately 30% by weight, along with any other contaminants. Continued operation may be hazardous.
- Low Level Control (LLC): A safety to turn the heat off if the liquid level falls below a safe level.
- Refrigeration Control (RT): Regulates the temperature of the cooling coils. Typically set at 45-50 °F. It is not necessary to change this setting when changing solvents.
- Solvent Vapor Control (SVC): Turns off the heat in the event of high vapor level and must be set at 10 °F below the boiling point of the solvent being used.
- Vapor Up Control (TH-1): Turns on an indicator light when the solvent vapors are up to proper level on the cooling coils. This indicates the equipment is ready to operate. It also inhibits the spray pump(s) when the vapor line is below the normal operating level. The proper temperature setting is 10 °F below the boiling point of the solvent being used.